DRX-RTD
RTD SIGNAL CONDITIONER

Made in the USA
INFINITY
Electromagentic Compatibility
Year 2000 Compliant
What You need:
ò Your choice of the signal conditioner
ò IBM PC or compatible
ò Windows 3.1 or higher, or Windows 95
ò RJ12 connector with telephone cable
ò Adapter (DB9/25 to RJ12)
SETTING UP YOUR SYSTEM
DRN Computer Connection

SPECIFICATIONS
100, 500, 1000 Ohm platinum RTD with both European (alpha=0.00385) and American (alpha=0.00392) curves.


DRX-RTDINPUT TYPE:

100 Ohm
-328 to 1562 ░F
-200 to 850 ░C
500 Ohm
-328 to 1562 ░F
-200 to 850 ░C
1000 Ohm
-328 to 1562 ░F
-200 to 850 ░C

ISOLATION

Three way isolation up to 1800 V Peak for 1 minute, 345 V Peak continuous per IEC spacing. Common mode rejection 100 dB

INPUT OVER-VOLTAGE PROTECTION
250 V AC for 1 Minute

ANALOG TO DIGITAL TECHNIQUE
Dual slope

READ RATE

3 readings/second, automatic polarity

ACCURACY AT 25 ░C

▒0.5 ░C for platinum RTDÆs

TEMPERATURE STABILITY
0.05 ░C/░C

STEP RESPONSE FOR RS485 OUTPUT

2 seconds to 99% of the final value (Filter time constant = 64)

RESPONSE TIME

To verify the response time, check the carriage return <CR>, it will be sent at the end of the response. You can send another command after you receive the <CR>.
i.e. send: *01X01 / response: 01X01<DATA><CR>
Note: 01 is address.

Signal Conditioner
WARM UP TO RATED ACCURACY
30 minutes

INPUT POWER

10 to 32 Volt

DC POWER CONSUMPTION

2.4 Watts (100 mA at 24 V DC)

OPERATING AMBIENT

-5 to +55 ░C

STORAGE TEMPERATURE

-40 to +85 ░C

RELATIVE HUMIDITY

90% at 40 ░C non-condensing

Help Price List Specifications Print Manual Manuals Home